GE Xcellerex XDM Series Operating Instructions Manual page 110

Table of Contents

Advertisement

4 User interface
4.4 Detail screens
4.4.5 Temperature screens
Part
5
6
7
1
110
Description
P-Term (proportional)
I-Term (integral)
D-Term (derivative)
CV HiLimit
CV LoLimit
SP HiLimit
SP LoLimit
For example, the remote setpoint range of the TCU is -10°C to 90°C. The CV = 0% sets the TCU
remote setpoint to -10°C. The CV = 100% sets the TCU remote setpoint to 90°C.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Function
User-defined PID tuning parameters.
User-defined limits for temperature PID con-
troller output (controlled variable, CV). The CV
limits (in %) specify the allowed temperature
1
range for the TCU.
Note:
The default CV limits are suitable to use with
a TCU working range 6°C to 65°C. The CV
limits should be adjusted according to the
actual operating range of the TCU.
User-defined limits for the input range of
temperature setpoint (SP) of disposable bag
contents.
Note:
The SP limits specify the allowed temperature
range for the contents of the disposable bag.

Advertisement