6 Preparation
6.4 Prepare and install the probes
6.4.1 Calibrate the pH probe
Step
3
4
5
6
7
8
160
Action
1 Select Calibrate and press ENTER (2).
2 Select Sensor 1 and press ENTER.
3 Select pH and press ENTER.
4 Select Buffer Cal and press ENTER.
5 Select Auto and press ENTER.
6 Select Start Auto Cal and press ENTER.
7 Rinse the pH probe with de-ionized water.
8 Put the probe into the pH 4 buffer. Stir approximately ten times and make
sure no air bubbles are present on the probe.
9 Press ENTER to start the calibration. When the calibration reading is
finished, the system displays a range of pH values.
10 Select the closest value to pH 4 from the list with the navigation keys
and press ENTER.
Result: The system displays Place the sensor in Buffer 2.
Rinse the pH probe with de-ionized water.
Calibrate the probe at pH 10 as follows:
1 Put the probe in the pH 10 buffer. Stir approximately ten times and make
sure no air bubbles are present on the probe.
2 Press ENTER to start the calibration. When the calibration reading is
finished, the system displays a range of pH values.
3 Select the closest value to pH 10 from the list with the navigation keys
and press ENTER.
Result: The pH probe is autocalibrated to pH 4 and pH 10. The slope and
offset values are displayed.
Make a note of the slope and offset value from the display.
Tip:
Consult the probe manufacturer's manual for acceptable slope value range.
Press EXIT (3) repeatedly until you reach to the main screen.
Rinse the pH probe with de-ionized water.
Disconnect the probe cable from the pH probe.