6 Preparation
6.4 Prepare and install the probes
6.4.1 Calibrate the pH probe
Step
4
5
162
Action
Press MENU (1).
2
Result: A menu is displayed on the transmitter screen.
Use the navigation keys to make selections in the menu:
1 Select Calibrate and press ENTER (2).
2 Select Sensor 1 and press ENTER.
3 Select pH and press ENTER.
4 Select Standardize and press ENTER.
5 Enter the externally measured pH value of your sample and press ENTER.
6 Press EXIT (3) repeatedly until you reach to the main screen.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD