Connecting The Electrical Connections; Connecting Auxiliary Devices - Dräger Primus SW 4.5n Instructions For Use Manual

Anesthesia workstation
Table of Contents

Advertisement

Assembly and preparation

Connecting the electrical connections

Connecting auxiliary devices

The Primus has two auxiliary outlets on the rear of
the machine. Each outlet is rated 4 amps and pro-
tected by safety fuses.
A
1 Connect to auxiliary outlets (A) at the back of
the workstation.
Install the Desflurane vaporizer in its mount and
connect it to the auxiliary outlet at the back of
the machine, see page 77.
CAUTION
Risk of device failure
In the event of a power failure, any devices con-
nected to auxiliary power outlets will not be pow-
ered by the UPS.
Pay special attention to all power indicators of
connected devices.
68
CAUTION
Risk of device failure
If HF surgical devices are connected to the auxil-
iary outlets, the leakage current may influence the
electronics of the anesthesia machine causing it
to fail.
Do not connect HF surgical equipment to the an-
esthesia machine's auxiliary outlets.
Make sure the maximum power consumption of the
auxiliary systems does not exceed permissible val-
ues: See the corresponding instructions for use.
WARNING
Risk of electric shock
Connecting devices to the auxiliary outlets of
the anesthesia machine can cause an in-
crease in leakage current beyond permissible
values if the protective conductor of a device
fails.
Check the leakage current when connecting
devices to the auxiliary outlets. If connecting
a device or several devices increases the leak-
age current to a value which exceeds the per-
missible value, do not use the auxiliary outlets
of the anesthesia machine. Instead, use a
mains power socket on the wall.
The system must fulfill the requirements for
medical electrical equipment in accordance
with the relevant standards, see "Technical
Data", page 261.