Monitoring
Test Considerations and Oximeter
Accuracy
Functional Testers and Patient Simulators
For functional testing of the pulse oximeter sensors
and cables with a functional tester or patient simu-
lator follow the individual testing device's operator's
manual, especially regarding the suitability and ac-
curacy of the simulated values.
CAUTION
Risk of inaccurate data
If simulators are used as calibrators, the SpO
module may produce incorrect data.
Simulators must not be used as a calibrator.
160