Dräger Primus SW 4.5n Instructions For Use Manual page 249

Anesthesia workstation
Table of Contents

Advertisement

1)
Cylinder pressure
Measuring range
Resolution of the displayed value
Accuracy
Flow measurement (hot-wire anemometry)
All values apply under ATPD conditions:
Flow
Measuring range
Resolution of the measurement
Accuracy at 60 L/min
Tidal volume V
T
Measuring range
Resolution of the displayed value
Accuracy
Delta V
T
Measuring range
Resolution of the displayed value
Accuracy
Volume VT
INSP
Measuring range
Resolution of the displayed value
Accuracy
Minute volume MV
Measuring range
Resolution of the displayed value
Accuracy
1) Valid for Silverline pressure regulators.
Instructions for use Primus SW 4.5n
0 to 250 kPa x 100 (0 to 3600 psi)
1 kPa x 100 (14 psi)
±4 % or ±6 kPa x 100 (±4 % or ±87 psi)
–180 to 180 L per minute
0.1 L per minute
±8 % of the measured value
0 to 9999 mL
1 mL
±8 % of the measured value or ±5 mL,
the greater value applies
0 to 9999 mL
1 mL
±16 % or ±10 mL, the greater value applies
0 to 9999 mL
1 mL
±8 % of measured value or ±5 mL, the greater value applies
0 to 99.9 L per minute
0.1 L per minute
±8% of measured value or ±0.1 L/min, the greater value ap-
plies
Technical Data
249