Cleaning, Disinfection and Sterilization
Items which can be re-
processed
Primus workstation
Power cable, com-
pressed gas hoses, po-
tential equalization ca-
ble
Breathing hoses
Y-piece
Breathing bag with con-
nector and hose
Flexible arm for breath-
ing bag
(optional)
1)
Ventilator diaphragm
Cover of breathing sys-
tem with APL valve
Middle and bottom part
of breathing system
Expiratory port/inspira-
tory port
Absorber and insert
Spirolog/SpiroLife flow
sensors
to scavenger
1) Drain any water which may have collected in the ventilator diaphragm.
Larger quantities of condensation may impair operation of the workstation and/or lead to failure of the equipment!
2) The valve plate must be sterilized after washing in order to dry it. Correct operation of the workstation may be impaired and
lead to failure of the workstation if the control areas in the valve plate are not dried completely.
220
Recommended re-
Preclean-
processing inter-
vals
Daily
Monthly
Daily
yes
Daily
yes
Daily
yes
Weekly
yes
Weekly
Weekly
yes
Weekly
yes
Weekly
yes
Weekly
yes
Note the instructions for use of the flow sensors
Note the instructions for use of the AGS
Machine
ing
cleaning
and disin-
fection
no
no
no
no
yes
ing instructions for use.
yes
yes
ing instructions for use.
yes
no
yes
yes
yes
yes
yes
Manual
Cleaning
Disinfec-
tion
Outside
no
yes
no
Observe correspond-
no
yes
Observe correspond-
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
Instructions for use Primus SW 4.5n