Ventilator - Dräger Primus SW 4.5n Instructions For Use Manual

Anesthesia workstation
Table of Contents

Advertisement

Technical Data
O
-Flush
2
O
safety flow
2
Accuracy
Optional outlets
Air outlet
O
outlet
2

Ventilator

Electronically controlled, electrically driven piston ventilator, fresh gas decoupled
Ventilation modes
Settings
Pressure limitation P
In Volume Mode and
Volume AF Mode
Accuracy
Inspiration pressure P
In Pressure Mode
Accuracy
242
>35 L per minute
Standardized to ATPD conditions.
0 to 12 L per minute
Standardized to STPD conditions.
±40 % or ±2 L per minute, the greater value applies
For connecting O
Pressure corresponds to inlet pressure.
The outlets are not pressure monitored; this monitoring must
be ensured by the connected device.
Maximum permitted flow: 70 L per minute
Standardized to STPD conditions.
Maximum permitted flow: 20 L per minute
Standardized to STPD conditions.
Volume Mode, Pressure Mode, Volume AF Mode (op-
tional), CPAP (optional) and Pressure Support (optional)
MAX
(PEEP +10) to 70 hPa [(PEEP +10) to 70 cmH
±10 % of set value or at least ±3 hPa (±3 cmH
INSP
(PEEP +5) to 70 hPa [(PEEP +5) to 70 cmH
±10 % of set value or at least ±3 hPa (±3 cmH
flow meters and endotracheal suction.
2
Instructions for use Primus SW 4.5n
O]
2
O)
2
O]
2
O)
2