Technical Data
Compliance C
PAT
Measuring range
Resolution of the displayed value
Accuracy
MV
LEAK
Measuring range
Resolution of the displayed value
Accuracy
MV
MAND
Measuring range
Resolution of the displayed value
Accuracy
MV
SPON
Measuring range
Resolution of the displayed value
Accuracy
O
Uptake
2
Measuring range
Resolution of the displayed value
Accuracy
MV*CO
2
Measuring range
Resolution of the displayed value
Accuracy
250
0 to 250 mL/hPa (0 to 250 mL/cmH
0.1 mL/hPa (0.1 mL/cmH
±15 % of the measured value or ±0.5 mL/hPa (±0.5 mL/
cmH
O), the greater value applies
2
0 to 9.99 L per minute
0.01 L per minute
±15 % of (MV
+ MV
EXP
the greater value applies
0 to 99.9 L per minute
0.1 L per minute
±8 % of measured value or ±0.1 L per minute, the greater
value applies
0 to 99.9 L per minute
0.1 L per minute
±8 % of measured value or ±0.1 L per minute, the greater
value applies
0 to 9999 mL per minute
1 mL per minute
±15 % or ±20 mL per minute, the greater value applies
0 to 9999 mL per minute
1 mL per minute
±20 % or ±20 mL per minute, the greater value applies
O)
2
O)
2
) or ±0.1 L per minute,
LEAK
Instructions for use Primus SW 4.5n