Start Manual Ventilation/Spontaneous; Breathing - Dräger Primus SW 4.5n Instructions For Use Manual

Anesthesia workstation
Table of Contents

Advertisement

Spontaneous breathing
C
Set the lever (C) of the APL pressure control
valve to SPONT.
The pressure control valve is now open for free
spontaneous breathing, regardless of the set pres-
sure limit.

Start manual ventilation/spontaneous

breathing

A
1 Press the Man.Spont. key (A); its LED and the
status line flash.
2 Confirm via rotary knob (B).
Instructions for use Primus SW 4.5n
Man.Spont.
Certain alarms are disabled automatically in the
ventilation mode Man.Spont. in order to avoid arti-
facts.
See page 128 for a list of alarms which are active in
the Man.Spont. mode.
B
Operation
97