Cleaning, Disinfection and Sterilization
Removing the anesthetic gas
scavenging system AGS
B
1 Remove the scavenging hose (A) from the AGS
system on the back of Primus.
2 Remove the gray transfer hose (B).
3 Remove the anesthetic gas scavenging system
(C).
4 Prepare the individual parts for conditioning in a
cleaning and disinfecting machine.
Flow tubes must not be placed in the washer!
5 Remove the container for buffer volume.
214
B
C
A
6 Unscrew the union nut (D).
7 Unscrew the flow tube (E).
8 Unscrew the union nut and remove the particle
filter (F).
The particle filter may be disposed of with ordi-
nary domestic waste after being sealed, see
"Maintenance" on page 229.
Note the instructions for use of the anesthetic gas
scavenging system AGS.
Instructions for use Primus SW 4.5n
D
E
F