Removing the secretion aspiration sys-
tem (optional)
E
A
B
C
D D
1 Remove the suction hose (A) and vacuum hose
(B) leading to the endotracheal aspiration sys-
tem.
2 Grip the silicone sleeve (C) of the bottle cap
(see details below) and pull it off.
3 Remove the secretion collecting bottle (D) and
rinsing bottle from the holder and empty them.
Note the hygiene regulations of the hospital.
Both secretion collecting bottles may contain in-
fectious secretions.
NOTE
Always wear gloves when emptying the bottles.
4 Remove the filter (E) at the bottom of the en-
dotracheal aspiration system.
(See "Maintenance" on page 229.)
Instructions for use Primus SW 4.5n
Cleaning, Disinfection and Sterilization
5 Remove the silicone sleeve (F) from the lid (G)
of the secretion collecting bottle.
6 Remove the float (H) of the overflow protection
from the rising pipe.
Prepare associated parts for conditioning in a
cleaning and disinfecting machine so that they
can subsequently be relocated without difficulty.
When using the disposable VacuSmart container:
Dispose of the VacuSmart container and hose with
the infectious secretion.
Note the instructions for use of the suction unit.
The system has been designed and verified for the
procedures, chemicals, and values given in this
chapter. If other procedures, chemicals, and/or val-
ues are used, the user assumes all responsibility.
G
H
F
215