Monitoring

Mini trends (optional)

There are 3 different mini trends available that can
be displayed below the waveform area:
– MV*CO
2
– O
uptake
2
– C
/PEEP
PAT
To configure a mini trend:
Press the soft key screen layout (A).
Select a parameter module and select the corre-
sponding mini trend.
150
Mini trend for MV*CO
MV*CO
2
A
800
400
0
This mini trend displays the expiratory minute vol-
ume in combination with expiratory CO
tion over 15 minutes.
The current value of MV*CO
meric value above the mini trend.
The parameter MV*CO
that is expired by the patient.
The scaling depends on the expiratory tidal volume
V
and is automatically adjusted.
T
V
T
mL
50
150
500
1000
2
5
is displayed as nu-
2
indicates the CO
2
MV*CO
2
50
150
500
1000
Instructions for use Primus SW 4.5n
mL/min
concentra-
2
volume
2