Dräger Primus SW 4.5n Instructions For Use Manual page 265

Anesthesia workstation
Table of Contents

Advertisement

Immunity against
IEC 60601-1-2 test
level
RFcoupled into
150 kHz to 80 MHz:
lines
10 V inside ISM
(IEC 61000-4-6)
bands
150 kHz to 80 MHz:
3 V outside ISM
bands
1) For P
, insert the highest possible "equivalent isotropic radiated power" of the adjacent radio fre-
EIRP
quency transmitter. In the vicinity of equipment marked with the symbol
Field strengths from fixed, portable, or mobile radio frequency transmitters at the location of the med-
ical device should be less than 3 V/m in the frequency range from 150 kHz to 2.5 GHz and less than
1 V/m above 2.5 GHz.
2) ISM bands in this frequency range are: 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz;
26.957 MHz to 27.283 MHz; 40.66 MHz to 40.70 MHz.
Instructions for use Primus SW 4.5n
Compliance
level (medical
device)
10 V
2)
3 V
2)
Electromagnetic environment
Recommended minimum distance to
portable and mobile radio frequency
transmitters with transmission power
P
to the medical device including its
EIRP
lines:
1)
(1.84 m x √P
)
EIRP
1)
(6.04 ft x √P
/W)
EIRP
, interference can occur.
Technical Data
265