Damage Prevention - Ametek ta7000R User Manual

Trace analytical gas purity monitor
Hide thumbs Also See for ta7000R:
Table of Contents

Advertisement

4 each, required for use with VICI-VCR adapters at regulators.
Nut and ferrule, 1/8-inch VICI
1 each required for use in connecting the Actuator gas.
Gasket, 1/4-inch VCR, Nickel
8 each, required for use with VICI-VCR adapters and purifier connections.
Tubing, 1/16-inch OD x 0.03-inch ID, Stainless Steel
4 each approximately 72" lengths for gas supply umbilical.
Tubing, 1/8-inch OD, for pneumatic (Actuator gas) line
1 each approximately 72" length for gas supply umbilical.
Guidelines for prevention of damage
!
CAUTION
Carrier gas
The RGD detectors can be destroyed if operated at temperature without
flow. The carrier gas should remain flowing until all heated zones are
below 50o C (2 to 3 hours).
Line voltage
Do not connect the power cord to line voltage without checking the cor-
rect operating voltage on the label located on the rear of the instrument.
Actuator Gas pressure
Always maintain correct actuator gas pressure. Low pressure (below 70
psig) can result in failure of the internal valves and /or stoppage of the gas
flow to sensitive instrument components. High pressure (above 150 psig)
will damage the valve actuators.
Shut down
Be sure all heated zones are shut off prior to disconnecting gas supply
lines. The carrier gas should remain flowing until all heated zones are
below 50o C (2 to 3 hours). After shutting off carrier gas, immediately cap
all open rear panel gas ports.
Electrical signal cables
Turn power off before connecting or disconnecting electrical signal cables.
Damage to circuit boards could result.
PLEASE STUDY THIS MANUAL THOROUGHLY BEFORE IN-
STALLING OR OPERATING THE INSTRUMENT.
User Manual | 27