Unpacking and Inspecting
• Inspect the shipping container exterior for evidence of damage or indications
• Carefully open the shipping container and remove the analyzer. Inspect
• Do not discard the shipping container until installation is complete.
• Analyzer
• Power Cord
• Operating Manual

Location and Environment

24 | Model ta7000R Gas Purity Monitor

INSTALLATION

In order to have a smooth installation of the Trace Analytical Reduction
Gas Analyzer and to get the maximum performance, several aspects
should be considered. This guide clarifies the important points for a suc-
cessful installation.
Perform the following inspection to assure that the analyzer has not been
damaged during shipment.:
of excess moisture before unpacking.
the exterior of the analyzer for noticeable damage.
Upon receipt of the analyzer, check the interior of the shipping container
for the following items:
If any evidence of damage exists, immediately contact the shipping com-
pany, save the original shipping container, and notify Molecular Analytics
as soon as possible for consultation.
1.
The area where the Reduction Gas Analyzer is to be operated
should be clean with minimum particulate matter and dust in the air. In-
stall the instrument in an area where there are no vibrations to affect the
electronics and detectors. There should not be any draft directly on the