Error Messages

Chassis Controller Error
Detector Controller Error
Temperature Out of Range
Vsig Out of Range (Lamp OFF ?)
Carrier Pressure Out of Range
46 | Model ta7000R Gas Purity Monitor
ERROR MESSAGES
Error Messages are viewed and acknowledged on the System Settings
screen. When the Status LED on the front panel is red this indicates that
there are system errors or the status of the analyzer has changed. The
System Status page displays error messages that the system has encoun-
tered. See the Run section for more information on the System Settings
page.
Internal electronics failure - Call manufacturer or qualified service repre-
sentative.
Internal electronics failure - Call manufacturer or qualified service repre-
sentative.
Either the temperature zone is 2 degrees over the set point, or the tem-
perature zone has been disabled. Perform the Warmup procedure to clear
this error.
During the Warmup cycle the analyzer performs a system test of the RGD
lamp. If the RGD lamp is not within operating range this error will be
displayed.
External carrier pressure is not within specified limits. Verify external car-
rier pressure supply.