Com Ports - Ametek ta7000R User Manual

Trace analytical gas purity monitor
Hide thumbs Also See for ta7000R:
Table of Contents

Advertisement

The following defi nes each fi eld:
DDDD-DD-DD TT:TT:TT,#,R,S,A,EEEEEEEE,NNNNN,C, ...
DDDD-DD-DD
TT:TT:TT
#
R
A
S
EEEEEEEE
ror)
1.
General Alarm (MSD)
2.
Spare
3.
Spare
4.
Spare
5.
Spare
6.
Spare
7.
Spare
8.
Spare (LSD)
NNNNN
C
The following is an example of this output string:
2001-08-16 14:35:00,99,R,S,1,10000000,H2 ,0,CO ,2483, , , , , ,
Com Ports - COM 1, COM 2, and COM 3
All three of these ports are 9-pin male D-type connectors. Pin 2 is used
to transmit data to other equipment. The analyzer uses Pin 3 to receive
data. Pin 5 is a common ground. All other pins are unused.

COM PORTS

= Date (2000-03-23)
= Time (12:23:44)
= Analyzer Serial Number
= Run or Maintenance (Calibration) Mode
= EDL or Single Analysis Mode
= N/A
= Eight digits of error status (0 = Not Error, 1 = Er-
= Name of impurity
= Concentration in ppb x 100
User Manual | 19