Ametek ta7000R User Manual page 34

Trace analytical gas purity monitor
Hide thumbs Also See for ta7000R:
Table of Contents

Advertisement

Warmup Mode Screen
Warmup Pass Screen
34 | Model ta7000R Gas Purity Monitor
After pressing Warmup the instrument will begin heating critical zones
and will perform a series of system tests. The results from these tests will
appear on the Warmup screen. The Warmup menu item will also change
to Cool. If you wish to stop the warm-up process, press the Cool key and
the instrument will return to a completely neutral state. If any errors oc-
cur during the warm-up process, proceed to Appendix B – error messages.
Warmup Mode
Module Electronics:
Detector System:
Detector Temperatures:
Carrier Pressure:
Cool
The Warm-up process typically takes 20 to 30 minutes. When the warm-
up process has completed succesfully a Run munu label will appear.
Pressing the Run key will take the user to the main screen where the Main
menu will be available.
Idle Mode
Module Electronics:
Detector System:
Detector Temperatures:
Carrier Pressure:
***** Successful Warm-up *****
Run
Cool
12:35:25
WARM-UP
Pass
WARM-UP
Pass
Pass
Pass
Pass
Page
12:34:25
Page