Sample Gas Supplies

Gas Ports
Sample Gas
16 | Model ta7000R Gas Purity Monitor
Inlet Pressure Range
Inlet Pressure Stability
Return Pressure
Flow Rate
Temperature
Maximum Impurity Levels
Carrier/Sample Inlet
Aux (Calibration)
Sample Outlet
Actuator Gas
Inlet Fitting
Inlet Pressure Range
Inlet pressure stability
Vent pressure
FlowRate
70 to 90 psig (4.8 to 6.1 bar)
+ 2%, regulator required
Atmospheric vent is optimal, + 0.5 psig
maximum
300 cc/min minimum, sample bypass at 50
cc/min.
60 to 100 °F (16 to 38 °C), optimum when
maintained + 2°C
200 ppb Moisture, CO2, Hydrocarbons
1/4-inch VCR compatible
1/16-inch VICI compression
1/16-inch VICI compression
1/8-inch VICI compression
1/4-inch Male face seal fitting*
70 to 90 psig (4.8 to 6.1 bar)
+ 2%, UHP regulator required
Atmospheric pressure vent is opti
mal, + 0.5 psig maximum
300 cc/min minimum, sample bypass
at 50 cc/min