IEEE-Common Commands

Command:
*CLS
Syntax:
*CLS
Description:
The Clear Status (*CLS) command clears the status byte by emptying the error queue and clearing all the event registers
(SESR) including the Data Questionable Event Register, the Standard Event Status Register, the Standard Operation Status
Register and any other registers that are summarized in the status byte.
Parameters:
none
Response:
none
Example:
*CLS
Command:
*ESE
Syntax:
*ESE<wsp><value>
0
Description:
The standard Event Status Enable command (*ESE) sets bits in the Standard Event Status Enable Mask (SESEM) that enable
the corresponding bits in the standard event status register (SESR).
The register is cleared:
at power-on,
by sending a value of zero.
The register is not changed by the *CLS command.
Parameters:
The bit value for the register (a 16-bit signed integer value):
Bit
7 (MSB)
6
5
4
3
2
28
Common Commands
IEEE-common commands which can be used with the instruments. This
section gives fuller descriptions of each of these commands.
value
255
Mnemonic
Power On
Not Used
Command Error
Execution Error
Device Dependent Error
Query Error
on page 15 gave a brief introduction to the
Keysight N778xC Series Polarization Instruments Programming Guide
Decimal Value
128
64
32
16
8
4