Keysight Technologies N778 C Series Programming Manual page 66

Polarization instruments
Table of Contents

Advertisement

5
Polarization Commands
Command:
:POLarimeter:STOP
Syntax:
:POLarimeter:STOP
Description:
Stops logging
Parameters:
None
Example:
:pol:swe:stop
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:ZERO
Syntax:
:POLarimeter:ZERO
Description:
Zeros photodiodes. That means dark current is measured and will be subtracted from future measurements.
Parameters:
None
Example:
:pol:zero
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:ZERO?
Syntax:
:POLarimeter:ZERO?
Description:
Returns zero result
Parameters:
None
Response
0: last zero was successfull
1: last zero failed (too much light, for example)
Example:
:pol:zero?
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:LOOP
Syntax:
:POLarimeter:SWEep:LOOP
Description:
Sets the number of loops, which should be logged after starting
66
Keysight N778xC Series Polarization Instruments Programming Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7781cN7785cN7786cN7788c

Table of Contents