Keysight Technologies N778 C Series Programming Manual page 67

Polarization instruments
Table of Contents

Advertisement

Parameters:
Integer
0: endless
Example:
:pol:swe:loop 1
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:LOOP?
Syntax:
:POLarimeter:SWEep:LOOP?
Description:
Gets the number of loops, which should be logged after starting
Parameters:
None
Response:
Integer
Example:
:pol:swe:loop?
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:STARt
Syntax:
:POLarimeter:SWEep:STARt<wsp>[SOP|SOPCONTINUOUS]
Description:
This command starts the polarimeter logging.
Parameters:
None: Starts logging as configured
SOP: Starts logging and sets loop to 1
SOPCONTINUOUS: Starts logging and sets loop to 0 (endless logging)
Example:
:pol:swe:star sop
Affects
N7781C / N7786C / N7788C
Command:
POLarimeter:SWEep:STATe?
Syntax:
POLarimeter:SWEep:STATe?
Description:
This command returns the current logging state
Return value status string e.g SAMPLING,NO_DATA
Parameters:
None
Keysight N778xC Series Polarization Instruments Programming Guide
Polarization Commands
5
67

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7781cN7785cN7786cN7788c

Table of Contents