Parameters:
none
Response:
Selftest failed 1
A value of zero indicates no errors.
Example:
*TST? > 0
Command:
*WAI
Syntax:
*WAI
Description:
The WAIt command prevents the instrument from executing any further commands until the current command has finished
executing. Some module firmware includes commands that set a "StatNOPC" flag during execution to indicate that the
module is busy. *WAI blocks the GPIB bus to all commands until every module hosted by the instrument is no longer busy.
All pending operations, are completed during the wait period.
Parameters:
none
Response:
none
Example:
*WAI