Keysight Technologies N778 C Series Programming Manual page 69

Polarization instruments
Table of Contents

Advertisement

Parameters:
None
Response:
Integer
Example:
:pol:swe:get:ind? → +1120055703
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:SAMPles
Syntax:
:POLarimeter:SWEep:SAMPles
Description:
This command sets the number of samples/logging count (1-1048576)
Parameters:
Number of samples
Example:
:pol:swe:samp 1000
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:SAMPles?
Syntax:
:POLarimeter:SWEep:SAMPles?
Description:
This command gets the number of samples/logging count
Parameters:
None
Response:
Interger
Example:
:pol:swe:samp? → +1000
Affects
N7781C / N7786C / N7788C
Command:
:POLarimeter:SWEep:SAMPles:CURRent?
Syntax:
:POLarimeter:SWEep:SAMPles:CURRent?
Description:
This command gets the number of already logged values. Will be 0, if logging finished.
Parameters:
None
Response:
Number of already logged samples.
Example:
:pol:swe:samp:curr? → +34
Affects
N7781C / N7786C / N7788C
Keysight N778xC Series Polarization Instruments Programming Guide
Polarization Commands
5
69

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7781cN7785cN7786cN7788c

Table of Contents