Command:
:PCONtroller:GEN:SCRAmble?
Syntax:
:PCONtroller:GEN:SCRAmble?<wsp><number>
Description:
Generate a random sequence.
Example: :PCON:GEN:SCRA? #Datapoints
Example: :PCON:GEN:SCRA 120000
Return Value = status string "GEN DONE"
Parameters:
<number> integer value from 1-1000000
Response:
Status message: "GEN DONE" when the generation is finished
Example:
:PCONtroller:GEN:SCRAmble? 300' -> "GEN DONE"
Affects
N7785C / N7786C / N7788C
Command:
:PCONtroller:SWITch
Syntax:
:PCONtroller:SWITch
Description:
Sets the optical switch in the N7788C
Parameters:
Integer value
0 = optical Scrambler Input signal will be routed to the internal polarimerer
1 = optical Scrambler Input signal will be routed to the Scrambler Output
Response:
None
Example:
:pcon:swit 1
Affects
N7788C
Command:
:PCONtroller:SWITch?
Syntax:
:PCONtroller:SWITch?
Description:
Gets the optical switch build in the N7788C
NOTE: this value is the last set value to set the latching MEMS-switch!
Parameters:
None
Keysight N778xC Series Polarization Instruments Programming Guide