Troubleshooting: Camera Cart
3.5
Troubleshooting: Camera Cart
3.5.1
Possible Camera Occurrences
Camera Occurrences
There are several possible occurrences that can be attributed to communication failures related to
the IR camera.
Occurrence
Decrease in cam-
No communication
Not tracking mark-
er spheres proper-
Curve Dual Dis-
play only: Camera
motor control does
NOTE: If you are having problems with your camera, contact Brainlab support.
38
Symptoms
• Restricted field of view
for navigation
era volume
• LEDs in camera lens are
not on
Camera does not beep
No power
during boot-up
Camera does not beep
during application start up
Software displays uniden-
tifiable gray marker
spheres in the camera
ly
field of view
Pressing the motorized
camera control buttons
does not move camera
not work
Technical User Guide Rev. 1.2 Curve Dual Display, Curve Single Display Ver. 1.0
Possible Causes
• Overheating
• Communication error between the Monitor
and Camera Carts
• IR light from another device causing interfer-
ence (e.g., pulse oximeter).
• Surfaces in the OR are unintentionally re-
flecting IR light (e.g., reflective strips on
clothing)
• Problem with internal cabling
• Problem with external cabling
• If using wireless: Problem with power supply
• Communication error between Monitor and
Camera Carts: Check cabling/wireless con-
nection
• IR light from another device causing interfer-
ence (e.g., pulse oximeter).
• Surfaces in the OR are unintentionally re-
flecting IR light (e.g., reflective strips on
clothing)
• Motor on/off switch is turned off: Turn switch
ON
• Communication error: Check cabling/ wire-
less connection
• If green camera status LED is off, communi-
cation is not possible: Reboot device