Ups; Overview - Brainlab CURVE DUAL DISPLAY Technical User Manual

Hide thumbs Also See for CURVE DUAL DISPLAY:
Table of Contents

Advertisement

UPS

2.4
UPS
2.4.1

Overview

General Information
The Monitor Cart is equipped with an uninterruptible power supply (UPS) which provides power
to the system during a power failure.
NOTE: It is not possible to turn off the UPS. Contact Brainlab support if UPS deactivation is
required.
Powered Components
If mains power is lost while using the Curve system with wired camera communication, the
complete Monitor and Camera Carts are powered for a maximum of five minutes (when battery
is fully charged).
If mains power is lost while using the Curve Dual Display System with wireless camera
communication, only the Monitor Cart is powered. The Camera Cart loses power immediately
and requires up to two minutes to warm up after regaining mains power.
In case of power failure while charging, battery may not run for the full five minutes.
28
Technical User Guide Rev. 1.2 Curve Dual Display, Curve Single Display Ver. 1.0