ELECTRICAL SAFETY
4
ELECTRICAL SAFETY
4.1
Introduction
4.1.1
Monitor Cart
Monitor Cart Classification
The Monitor Cart is classified as Class I Equipment according to IEC 60601-1 and must be tested
accordingly.
Classification
Safety Requirements
To avoid the risk of electrical shock, the Monitor Cart must only be connected to a mains
power supply with protective earth.
The line connector fuse must have a 16A rating.
Technical User Guide Rev. 1.2 Curve Dual Display, Curve Single Display Ver. 1.0
Definition
Refers to equipment classification regarding protection against electric shock.
Class I
Protective means are provided for metallic accessible parts or metallic internal
parts, such as connection to PE (protective earth).