Troubleshooting: Camera Cart

3.5
Troubleshooting: Camera Cart
Camera Occurrences
There are several possible occurrences that can be attributed to communication failures related to
the IR camera.
Occurrence
Decrease in cam-
era volume
No power
No communication
Not tracking mark-
er spheres proper-
ly
Camera motor
control does not
work
NOTE: If you are having problems with your camera, contact Brainlab support.
42
Symptoms
• Restricted field of view
for navigation
• LEDs in camera lens
are not on
Camera does not beep
during boot-up
Camera does not beep
during application start up
Software displays uniden-
tifiable gray marker
spheres in the camera
field of view
Pressing the motorized
camera control buttons
does not move camera
Technical User Guide Rev. 1.1 Curve Dual Navigation Station Ver. 1.2
Possible Causes
• Overheating
• Communication error between the Monitor
and Camera Carts
• IR light from another device causing interfer-
ence (e.g., pulse oximeter).
• Surfaces in the OR are unintentionally reflect-
ing IR light (e.g., reflective strips on clothing)
• Problem with internal cabling
• Problem with external cabling
• Communication error between Monitor and
Camera Carts: Check cabling
• IR light from another device causing interfer-
ence (e.g., pulse oximeter).
• Surfaces in the OR are unintentionally reflect-
ing IR light (e.g., reflective strips on clothing)
• Motor on/off switch is turned off: Turn switch
ON
• Communication error: Check cabling
• If green camera status LED is off, communi-
cation is not possible: Reboot device