(4) Type in the IP address of the RU, "telnet 172.16.1.10" (default value).
"Vx works login" appears on screen.
(5) Type in "cr-ir346".
"Password:" appears on screen.
(6) Type in "cr-ir346".
"->" appears on screen.
(7) Type in "softupdate".
All the panel indicators on the RU are illuminated for about two minutes (with no
audible alarm).
REFERENCE
By performing "soft update", the application software and the configuration data of default
values are written into the flash ROM.
(8) Make sure that "Write OK" appears on screen.
010-051-05
02.20.2002 FM3328
[Type]
(6) [Type] cr-ir346
(7) [Type]
(8) [Check]
CR-IR346RU Service Manual
FR6H2567.EPS
FR6H3097.EPS
(5) [Type]
FR6H3098.EPS
MC - 278