(2) Determine the locations for embedding the anchor nuts.
98.5
244
(3) Drill holes for embedding the anchor nuts, and then embed the anchor nuts.
52mm
(4) Secure the table in place.
[Secure] Bolt 12x20 (x4)
010-051-01
010-051-00
08.30.2001 FM3058
10.10.2000 FM2887 (1)
525.2 (reference value)
Retainer
Marking location x4
18φ
CR-IR346RU Service Manual
Table leg
Unit: mm
FR6H7612.EPS
Anchor nut
Na12
Table
FR6H7615.EPS
IN - 34
#2
[Embed] Anchor nut
#1 [Attach]/#3 [Remove]
Bolt 12x20 (x4)
W12 (x4)
FR6H7723.EPS