[6-4-4] Subscanning grip operation checkout
CAUTION
With software version A02 or earlier, this feature is not implemented, so it should not be used.
Returns the grip mechanism of the subscanning unit to the HP.
The associated unit location and operating I/Os are shown below:
Driven-side grip release HP sensor
SZ3
Driving-side grip release HP sensor
SZ2
Grip drive motor
MZ2
Subscanning unit
Unit Z
I Procedure
(1) [6] [ENT] ¡ [4] [ENT] ¡ [4] [ENT] ¡ [1] [ENT]
The system then returns the grip mechanism of the subscanning unit to the HP and
the display reads "RESULT: OK".
I Display
MU>UNIT>SRG>1 ......................................... (1)
|
|
|
COMMAND IS IN PROGRESS
|
RESULT : OK
010-051-05
02.20.2002 FM3328
FRONT
M
M
CR-IR346RU Service Manual
FR6H3066.EPS
MU - 158
|
|
|
|