10. Final Placement
The following types of installation methods are available for installing the machine. Consult
with the user for his request, and then install the machine according to one of the following
methods.
• Securing the machine with the adjustable feet.
• Securing the machine with the retainers.
• Securing the table and machine together.
• Securing the machine with the retainers and table.
10.1 Securing the Machine with the Adjustable Feet
(1) Install the four adjustable feet down onto the floor.
N NOTES N
When securing the adjustable feet, pay attention to the following points.
• The four adjustable feet should be evenly loaded.
However, if there is too little space on the right and left sides in the machine installa-
tion place to secure the adjustable feet on the rear side, the two adjustable feet on
the front side should be secured in place. In this case, the two adjustable feet and
two casters should be evenly loaded.
• When the adjustable feet are secured, the casters should be detached from the floor.
(2) Secure the adjustable feet.
010-051-00
010-051-01
10.10.2000 FM2887 (1)
08.30.2001 FM3058
Caster
CR-IR346RU Service Manual
IN - 29
Machine
[Secure] Na12 (x4)
Adjustable foot
FR6H7724.EPS