8.
Cassette Set Unit
To gain access to the locations to be cleaned, the suction cup assembly should be removed.
8.1
Removing the Cassette Set Unit
(1) Remove the suction cup assembly.
CAUTION
When installing or removing the suction cup assembly, support the IP removal arm by
hand.
If the procedures are performed without supporting it, the IP removal arm may be bent
or otherwise damaged.
REFERENCE
Move the IP removal arm to the IP suction position before proceeding to the next step.
FRONT
PUSH
010-051-00
10.10.2000 FM2887 (1)
IP removal arm
#2 [Remove]
DT3x6 (x2)
PUSH
#1 [Remove] Lid opening bracket
CR-IR346RU Service Manual
Cassette set unit
FR6H6076.EPS
Suction cup assembly
IP removal arm