12. Image/Conveyance Checks
In this section, the following checks are made as last-time verification for final placement.
• Conveyance checks
• Image checks
I Preparation Procedures
N NOTES N
• For the IP used for image/conveyance checks, an IP of the maximum size used in the
hospital should be employed.
• Make sure that the IP to be used does not contain any important image data
exposed.
(1) Subject the IP to be used to primary erasure.
"RU Instruction Manual/2.4 When Only Image Erasure is to be Performed (Primary
Erasure)"
(2) Place a steel rule (150mm) on the cassette and expose it at 1 mR.
REFERENCE
If the IP cannot be exposed at 1 mR, it should be exposed at a level equivalent to 1
mR, with reference to the following exposure conditions.
Note, however, that those conditions are for standard IP setup reference.
Distance:
Voltage:
Amperage: 50 mA
Time:
010-051-00
10.10.2000 FM2887 (1)
Cassette
Steel rule
FR6H7009.EPS
1.8 m
80 kV
0.013 sec
CR-IR346RU Service Manual
IN - 37
IN - 37