3.
Preparation for Installation
3.1
Precautions Regarding Installation
The place where the machine is to be installed needs to meet the criteria set forth in "Getting
Started" and "I Installation Site Requirements".
Preinstallation procedures should be performed beforehand for necessary construction work,
electrical work, and air-conditioning system installation.
I Installation Site Requirements
Avoid the following installation sites:
• Places where the temperature changes drastically.
• Places near heat sources such as heaters.
• Places where water leakage or equipment submersion may occur.
• Places where corrosive gas may be generated.
• Dusty places.
• Places where the machine is subject to constant or excessive vibration or shock.
• Places that are exposed to direct sunlight.
I Handling Precautions for Printed Circuit Boards and Optical Unit
When servicing a board or optical unit, use a wristband to ground your body. If you body is
not grounded, the electrostatic buildup on your body may damage the electronic parts.