[6-3-2] Overall sensor monitoring
Monitors all the sensors simultaneously.
REFERENCE
This menu item monitors the sensor status that prevails when the monitoring command is
issued. To monitor sensor state transitions in real time, execute "[6-3-3] REAL-TIME MONI-
TOR ALL".
I Procedure
(1) [6] [ENT] ¡ [3] [ENT] ¡ [2] [ENT]
The system then presents the status information about all sensors and displays the
message "RESULT: OK".
I Display
MU>SNS>2 .............................................. (1)
|
|
Open:o/Close:x
|
———————————————––
|
SSSSSSSSSSSSSSLLL
|
AAAAABCCCCZZZZDDD
|
12345112342345SSS
|
|
|
————————————————–
|
ooxoxoxxooxoxooo
|
|
RESULT : OK
010-051-05
02.20.2002 FM3328
NNN
123
CR-IR346RU Service Manual
MU - 151
|
|
|
|
|
|
|
|
|
|
|
|