Hide thumbs Also See for FCR XG-1:
Table of Contents

Advertisement

2. Error Code Table
Error
Error Name
Code
12241
Scanner control board error (1)
12242
Scanner control board error (2)
12243
Scanner control board error (3)
12251
Photomultiplier control board error (1)
12252
Photomultiplier control board error (2)
12255
Leading-edge detection error
12256
Leading-edge detection timing error
12262
Polygon rotation error (2)
12263
Polygon rotation error (3)
12264
Polygonal rotation error (4)
010-051-06
04.20.2002 FM3386
During routine processing, error is detected for the PLL function, polygon rotation function, start-point interval function on the
SCN12A board.
During routine processing, because the PMT high-voltage value is not set correctly on the SCN12A board, error is detected
for the high-voltage command diagnostics.
During bootup self-diagnostics, error is detected for the start-point mask signal generated from the SCN12A board and the
index signal (PINDXL) from the polygonal mirror (POL).
During routine processing, only error for the PMT analog power supply signal (+15VOKH, +15VOKL) is detected (detail code:
008200). Note that if the HV is OFF, the detail code is 008000.
During routine processing, only error for the high-voltage power supply signal (HVOKH, HVOKL) is detected.
During bootup, only error for the leading-edge detection signal (SED1, SED2L) is detected.
During routine processing, IP conveyance overrun occurs due to mechanical error or SCN12A board control error, and only
error for the SEDTM and FCLKTM signals on the SCN12A board is detected.
• During routine processing, error for the signals (PONL, POKL, PINDXL) from the polygonal mirror (POL) and the start-point
interval function on the SCN12A board is detected (detail code: 003400).
• During routine processing, error for the polygon lock signal (PONL, POKL) and error for the start-point interval function and
polygon rotation detection function on the SCN12A board are detected (detail code: 003400).
During routine processing, only error for the lock signal (POKL, PONL) from the polygonal mirror (POL) is detected (detail
code: 002000). If error for the SEDTM or FCLKTM signal on the SCN12A board is detected, the detail code is 002003.
During routine processing, error for the signals (PONL, POKL, PINDXL) from the polygonal mirror and the SEDTM signal on
the SCN12A board is detected (detail code: 001002). Alternatively, error for the index signal (PINDXL) from the polygonal
mirror and the SEDTM and FCLKTM signals on the SCN12A board is detected (detail code: 0010030). Alternatively, error
for the signals (POKL, PINDXL) from the polygonal mirror and the SEDTM signal on the SCN12A board is detected (error
code: 001002).
CR-IR346RU Service Manual
Occurrence Condition
MT-25
Analysis
Flow
14.2
14.3
14.5
14.3
MT-99
MT-99
MT-98
MT-101
MT-100
MT-100
MT-100
[2.]
MT-25