8) Error Message
Error 01
▣ Reason
The weight is unstable to initialize the scale
Solution
☞
Put the scale on a stable place and turn on the power
Error 02
▣ Reason
Load cell connection failure or error in A/D conversion part.
☞
Solution
Check the load cell connector to see if the polarity of signal is reversed
Error 05
▣ Reason
You pressed any key for long time or problem of key part
Solution
☞
If there is no problem in key part, call your CAS dealer
Error 08
▣ Reason
You have set the operating condition of ZERO key or TARE key not to
operate when the indicator is not stable.
☞
Solution
Reset the operating condition of ZERO and TARE key in SET mode F09
Error 09
▣ Reason
Current weight deviates from zero range
Solution
☞
Set the operational range of the ZERO key within 2% or 10% of the
maximum capacity in SET mode F08
Error 10
▣ Reason
Tare weight exceeds the maximum capacity of the scale
☞
Solution
Tare weight should be lower than the maximum capacity.
Otherwise, the maximum capacity should be greater than the tare
Error 13
▣ Reason
The zero range deviates from the set range
☞
Solution
Confirm that there is nothing on the platter.
If nothing exists, do calibration in CAL mode