TEST 2
FUNCTION : Display test
KEY
ENTER: Go to menu
Selection mode
Other keys : Perform test
TEST 3
FUNCTION : A/D Conversion test
KEY
: Change gain.
: Change filter
ENTER: Go to menu
Selection mode
Note 1. Check whether digital value is changing. If the digital value is fixed or zero is
displayed, please check the connection of load cell.
TEST 4
FUNCTION : RS-232C test with computer (COM1)
KEY
: Transmit to PC
after increasing
value.
: Transmit to PC
after increasing
value
ENTER: Go to menu
Selection mode
Note 1. Do this test in Hyper Terminal of PC after the connecting serial port with PC.
- Port Selection in Hyper Terminal has the Flow Control item, you have to set nothing.
Note 2. Send no.1 in computer keyboard and check if indicator receives no.49
Send no.8 in indicator key( ) and check if computer receives no.8
Note 3. Baud rate should be specified in SET mode before testing.(F31)
DISPLAY
tESt
2
TEST 2 condition
8888888
TEST 2 is performed
DISPLAY
tESt
3
TEST 3 condition
Shows digital value of current weight.
97853
This value means converted digital value.
DISPLAY
tESt
4
TEST 4 condition
00------00
Wait for transmission and reception
03------00
Transmit : 3, Receive : none
08------49
Transmit : 8, Receive : 1
44
DESCRIPTION
DESCRIPTION
DESCRIPTION