13. OPTIONS

OP-1
■ Set F56 to 1 in set mode.
■ Specification
Output Current
Resolution
Temperature Coefficient
Maximum Load Resistor
■ When the display weight is "0", the output current is 0mA or 4mA.
When the display weight is maximum capacity of the indicator, the output current is
20mA or 24mA.

■ To use current to voltage

If you add a 250Ω shunt resistor, voltage output will be 1 - 5V or 0 – 5V or 0 – 6V.
(250Ω * 4mA ∼ 250Ω * 20mA)
+ 15V

Analog Output Interface (Current Output)

4 – 20mA, 0 – 20mA, 0 – 24mA
More than 1/2000
0.01%
500Ω MAX.
HI
LO
Shunt Resistor
67