CAL 5
FUNCTION : Span Calibration
KEY
ENTER :

Span calibration

Note 1. If span calibration is done without any error,
you will go to micro span calibration (CAL 6) automatically.
Note 2. If the span value is low,
error message "Err 24" is displayed. Calibrate with lower resolution.
Note 3. If the span value is high, error message "Err 25" is displayed.
CAL 6
FUNCTION : Check Micro Span Calibration
KEY
ENTER : Save and
go to next menu
Note 1. Confirm if the displayed weight is equal to the setting weight that you have
set in CAL 3 and remove the weight from the platter.
If "0" is displayed, set SW1 DIP1 to OFF then you will go to normal mode.
Note 2. The bias is "0" when the central lamp lights up as above display.
Each lamp means the bias of -0.3, -0.2, -0.1, 0, 0.1, 0.2, 0.3 from the left lamp
DISPLAY
Load the weight which was set in
CAL 5
CAL 3 and press ENTER key.
Analog value of
Span calibration.
load weight state
Span calibration is finished.
- - -
You will go to micro span
calibration(CAL 6) automatically.
DISPLAY
Setting weight is shown on the
500.0
display.
Check the setting weight and set
▽▽▽▼▽▽▽
SW1 DIP1 to off.
19