Error 25
▣ Reason
Span value is too high
Solution
☞
Load cell is damaged or setting of current resolution is not possible.
Calibrate with less resolution
Error 26
▣ Reason
Zero value is too high
☞
Solution
Check whether the platter is empty
Remove the setting cover and set SW2 DIP2 to on so that the zero value is
increased. Proceed calibration again after checking in test mode 3
Error 27
▣ Reason
Zero value is too low
Solution
☞
Check whether the platter is empty.
Remove the setting cover and set SW2 DIP1 to on so that the zero value is
decreased. Proceed calibration again after checking in test mode 3
Error 28
▣ Reason
The weight is unstable
Solution
☞
Check whether load cell is properly connected
22