ERROR MESSAGE AND TROUBLE SHOOTING

1. Error in Weighing Mode
Err 02
Reason: Failure in load cell connection or error in A/D conversion part.
Trouble shooting: Check the load cell connector so that you may see if
the polarity of signal is reversed.
Err 06
Reason: Error in printer connection
Trouble shooting: Check with printer connector.
If there is no problem with printer and printer connector,
please request A/S to head office.
Err 08
Reason: The ZERO key or TARE key is adjusted not to be operated under
the unstable condition.
Trouble shooting: Press ZERO or TARE key in stable condition.
Err 09
Reason: Current weight deviates from zero range.
Trouble shooting: Press the ZERO key within 10% of the maximum capacity.
Err 10
Reason: Tare weight exceeds the maximum capacity of the scale.
Trouble shooting: Set the tare to be smaller than the maximum capacity.
Otherwise the maximum capacity is reset to be larger than
the tare to be set in the calibration menu, and reset the
calibration using weight.
Err 13
Reason: The zero range deviates from the set range.
Trouble shooting: Confirm that there is nothing on the weighing platform.
If there were nothing, do calibration on CAL mode.
Over
Reason: The weight on platform is too heavy to be measured.
Trouble shooting: Do not load the item exceeds the maximum tolerance.
If the load cell is damaged, the load cell should be replaced.