Procedure - Keysight Technologies CX3300 Series User Manual

Current waveform analyzer 2- and 4-channel models
Table of Contents

Advertisement

Procedure

1. Set the slide switch on the test adapter to the 50 OHM. See
2. Connect the coaxial cable between the BNC jack connector on the test
3. Connect the CX1105A to the connection posts on the test adapter as shown in
4. Press the 1 (yellow), 2 (green), 3 (blue), or 4 (red) key associated with the
5. Press the Menu key several times to open the Configuration dialog box. The
6. Set the following conditions on the dialog box (Configuration > Calibration
7. Click the close (x) button located in the upper right corner on the Configuration
8. Press the Run key to start waveform acquisition.
9. Use the Vertical control knob to adjust the vertical scale. You will see square
Figure 1-6
Keysight CX3300 User's Guide, Edition 5
adapter and the Aux Out terminal on the CX3300 front panel.
Figure
1-6.
channel connected to the Aux Out terminal and enable the channel.
Menu key is located above the Horizontal control on the front panel.
Output).
Output State: ON
Source Shape: Square
The 5 V square pulse (1 kHz, 50% duty cycle) is applied from the Aux Out
terminal which has 50  output impedance.
dialog box to close the dialog box.
pulse (approximately 2.5 V, divided by two 50 ) as shown in
Test Adapter and Calibration Output
50 OHM
Slide switch
Connection posts
+
for connecting
the CX1105A
GND
-
BNC jack connector
Setting Up the CX3300
Verifying Basic Operation
Figure
1-6.
Figure
1-6.
27