Using The Cx1206A Sensor Head - Keysight Technologies CX3300 Series User Manual

Current waveform analyzer 2- and 4-channel models
Table of Contents

Advertisement

Performing User Calibration
Using the CX1101A/CX1102A/CX1103A

Using the CX1206A Sensor Head

The CX3300 cannot acquire the user calibration data of the CX1101A connected
to the CX1206A. Initially, the Gain value is set to 1 and the Offset value is set to 0.
If you want to use the user calibration capability, you need to obtain the Gain and
Offset values and enter the values to the CX3300 manually.
To obtain the user calibration data, you need a current source which supports 10
A output.
This section describes how to obtain the user calibration data by using the
following equipment and accessories.
Figure 3-2
90
Keysight B2900 (B2901A/B2902A/B2911A/B2912A) or equivalent
BNC-to-banana adapter, 2 ea.
BNC cable, 24 inches, 1ea., furnished with the CX3300
Connection Diagram
CX3300
CX1101A
CX1206A's
xtension cable
Current source
BNC-to-banana adapter
CX1206A's
high current adapter
Ground marker
CX1206A's
BNC-to-banana adapter
expander
Keysight CX3300 User's Guide, Edition 5
BNC cable