Keysight Technologies CX3300 Series User Manual page 34

Current waveform analyzer 2- and 4-channel models
Table of Contents

Advertisement

Setting Up the CX3300
Maintenance
Figure 1-7
34
CX3300/CX1100 Troubleshooting
If CX1100 is not detected by CX3300,
disconnect it from the analog input channel
and connect it to another channel.
Is CX1100 detected?
START
Yes
No
Perform
Fail
Self-Test.
Pass
Perform
Sensor Floating
Fail
PS test.
Pass
Go to CX1100
troubleshooting.
Keysight CX3300 User's Guide, Edition 5
CX3300 will be defective.
Contact nearest Keysight
Technologies sales office.