Keysight CX3300 User's Guide, Edition 5
Setting > Channels > Channel N screen:
Channel N: ON
•
Scale: 5 A/
•
Offset: 0 A
•
Bandwidth Limit: OFF
•
Invert Polarity: OFF
•
Setting > Sensor/Probe > Channel N screen:
Channel N: ON
•
Current Range: 10 A
•
Setting > Digital Channels screen:
Digital Channels: OFF
•
b. Control the current source (Keysight B2900) and apply the current of the
following condition. For setting up the current source, refer to its User's
Guide.
Condition:
Pulse base: 0 A
•
Pulse peak: 10 A
•
Pulse width: 1 ms
•
Duty cycle: 2.5% for the Keysight B2900 series
•
c. Control the CX3300 and acquire the waveform as shown below.
1. Press the Run/Stop key to start waveform acquisition.
2. Acquire the waveform as shown in
3. Press the Run/Stop key to stop waveform acquisition.
Performing User Calibration
Using the CX1101A/CX1102A/CX1103A
Figure
3-3.
93