Keysight Technologies CX3300 Series User Manual page 36

Current waveform analyzer 2- and 4-channel models
Table of Contents

Advertisement

Setting Up the CX3300
Maintenance
Figure 1-9
36
CX1152A Troubleshooting
Replace the grabber
of the failing channel with
a "working" grabber.
Does the problem
No
Replace the probe lead
of the failing channel with
a "working" probe lead.
Does the problem
No
a "working" flat cable
Replace the flat cable
with the "working" flat cable.
Does the problem
START
still exist?
Yes
still exist?
Yes
Do you have
No
No
for CX1152A?
Yes
Yes
No
still exist?
No
END
Keysight CX3300 User's Guide, Edition 5
Flat cable and/or pod
is defective.
Contact Keysight sales office
for further troubleshooting.
Pod is defective.
Contact Keysight sales office.